Skip to content

Secondary Menu


J.-P. Bouchard, I. Veilleux, R. Jedidi, I. Noiseux, M. Fortin, O. Mermut "Reference optical phantoms for diffuse optical spectroscopy. Part 1 – Error analysis of a time resolved transmittance characterization method" Opt. Express, 18 : 11495-11507, 2010

Abstract: Development, production quality control and calibration of optical tissue-mimicking phantoms require a convenient and robust characterization method with known absolute accuracy. We present a solid phantom characterization technique based on time resolved transmittance measurement of light through a relatively small phantom sample. The small size of the sample enables characterization of every material batch produced in a routine phantoms production. Time resolved transmittance data are pre-processed to correct for dark noise, sample thickness and instrument response function. Pre-processed data are then compared to a forward model based on the radiative transfer equation solved through Monte Carlo simulations accurately taking into account the finite geometry of the sample. The computational burden of the Monte-Carlo technique was alleviated by building a lookup table of pre-computed results and using interpolation to obtain modeled transmittance traces at intermediate values of the optical properties. Near perfect fit residuals are obtained with a fit window using all data above 1% of the maximum value of the time resolved transmittance trace. Absolute accuracy of the method is estimated through a thorough error analysis which takes into account the following contributions: measurement noise, system repeatability, instrument response function stability, sample thickness variation refractive index inaccuracy, time correlated single photon counting system time based inaccuracy and forward model inaccuracy. Two sigma absolute error estimates of 0.01 cm -1 (11.3%) and 0.67 cm -1 (6.8%) are obtained for the absorption coefficient and reduced scattering coefficient respectively. ***** This paper was published in Optics Express and is made available as an electronic reprint with the permission of OSA. The paper can be found at the following URL on the OSA website: http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-18-11-11495. Systematic or multiple reproduction or distribution to multiple locations via electronic or other means is prohibited and is subject to penalties under law *****

View at publisher

Secondary Menu


© 2008 INO.
All rights reserved.
Legal Notice